Introduction

The high-performance QES-PL photoluminescence quantum efficiency/spectral response testing system, independently developed by Oceanhood, is designed for measuring various materials such as solar cells and photodetectors. It is ideal for research and development as well as quality management in universities, research institutes, and enterprises.

Feature

◆ Due to the different spectral response ranges of detectors, the system provides a xenon lamp with high stability, exceeding 99%, ensuring high repeatability and accuracy of measurement results. The light source intensity can be automatically adjusted, and the deuterium lamp power is adjustable from 150W to 300W.

◆ The asymmetrical horizontal optical path with an achromatic design ensures symmetric spectral lines and good optical resolution.

Description

Item Value
Wide spectral measurement range 350nm-1700nm
Light source 300W high-stability xenon lamp
Dimensions 800mm W:700mmH:310mm / 800mm W:450mmH:415mm
Focal length 150mm
F/# F/4.5
Stray light 5×10-4
Minimum step size 0.005nm
Standard configuration gratings 1200 lines 500nm blaze grating, 600 lines 1000nm blaze grating
Slit effective size 10μm edge with automatic protection, width adjustable from 10μm to 3mm, height 5mm
Light source spectral linewidth ~0.5nm
Wavelength setting accuracy 0.1nm
Measurable parameters spectral responsivity, external quantum efficiency
Power meter: 190nm-2000nm, power range: 10μW-3W
Item Value
Quantum efficiency sample stage 1
Quantum efficiency light source 1
fiber 1
Digital source meter 1
Back electrode sample stage 1
Probe station 2
Probe wires 2
Probe box 1
SMA adapter 1
20X long focal length objective lens 1

More Info